Exposure Factors

65-70
Kilovoltage (kV)
20
Milliamperage (mAs)
Large Focus
Focal Spot Type
105-115 cm
Source-to-Image Distance

Equipment: With Bucky. Positions: Erect, semi-prone, or oblique.

Plate Size and Orientation

24 × 30 cm
Crosswise

Orientation: Plate in crosswise (landscape) position to include complete skull

Visible Anatomical Structures

Frontal Bone

Complete frontal region

Occipital Bone

Skull base

Sella Turcica

Dorsum and floor

Clinoid Processes

Anterior and posterior

Anatomical Reference Planes

CP - Coronal Plane

Passes through the EAM, aligned with the midline of the IR

MCP - Mid-Coronal Plane

Perpendicular to the table/Bucky

MSP - Mid-Sagittal Plane

Parallel to the plane of the IR

IOML - Infraorbitomeatal Line

Parallel to the transverse axis of the IR

Specific Centering Point

5 cm above the EAM

The central ray is centered 5 cm above the External Auditory Meatus

Reference: From the EAM upwards, midline of the IR

Patient Positioning

Patient in erect, semi-prone, or oblique position
Place the affected side against the Bucky or table
Head in a true lateral position
Coronal plane (CP) passing through the EAM aligned with the midline of the IR
MCP (mid-coronal plane) perpendicular to the table/Bucky
IPL (interpupillary line) perpendicular to the IR
MSP (mid-sagittal plane) parallel to the plane of the IR
IOML (infraorbitomeatal line) parallel to the transverse axis of the IR
Inion-acanthion and inion-nasion equidistant to the plate
Place hands on both sides of the head for support

Central Ray Direction

Perpendicular to the IR

PERPENDICULAR
5 cm above the EAM

Entrance point: 5 cm above the EAM of the opposite side

Exit point: 5 cm above the EAM of the supported side

Centering: IR centered with the central ray

No angulation - Direct 90°

Patient Instructions

"Hold your breath during the exposure"

Maintain complete immobility - Do not move head during exposure

Sella Turcica Evaluation

Normal Measurements

Length: 5-16 mm

Depth: 4-12 mm

Area: 30-130 mm²

Pathological Signs

Clinoid erosion

Increased size

Double floor of the sella

Technical Considerations

Symmetry

Ensure that inion-acanthion and inion-nasion are equidistant.

Rotation

Avoid rotation: MSP parallel to the IR, IPL perpendicular.

Key Structures

Focus on clear visualization of the sella turcica and clinoids.